Development of an Convenient and Robust Analytical Technique for HPLC-based Determination of Silodosin in Capsule Formulation
Md. Abid Hasan1,3*, Naima Helal1,2, Sajia Azmi3, Nursad Jahan Sultana3,
Sumaya Sarmina Nasir3, Rehnuma Jafreen4, Md Shadman Khan3, Zobayda Islam Juthi2,
Kazi Jannatun Nayeem3
1Department of Pharmacy, University of Asia Pacific, Dhaka, 1205, Bangladesh.
2Department of Pharmacy, World University of Bangladesh, Dhaka, 1230, Bangladesh.
3Department of Pharmacy, University of Science and Technology, Chittagong, Chattogram, 4202, Bangladesh.
4Department of Pharmacy, Jahangirnagar University, 1342, Bangladesh.
*Corresponding Author E-mail: mdabid780@gmail.com
ABSTRACT:
Benign prostatic hyperplasia is commonly treated with silodosin, a selective 1-adrenoreceptor antagonist. The goal is to create a straightforward and accurate high performance liquid chromatographic technique for silodosin detection and validate it in accordance with ICH standards. Agilent HPLC 1260 Infinity series was used and empower software was utilized for data processing. On a Cyano; 4.6mm × 25cm; 5mm, packing USP Octadecyl Silane column, silodosin was successfully separated by using mobile phase, made up of a mixer of 4.0 pH ammonium acetate buffer: Acetonitrile: Methanol (30:30:40) at 1.0ml/min flow rate and at a wavelength of 270 nm.The silodosin responded at 3.9minutes. The validation parameters including specificity, LOD/LOQ, linearity, accuracy, precision, robustness, and solution stability, were verified for performance of the method. All the peaks were well separated and there was no interference. Correlation coefficient of silodosin was 1.00, which indicated the method maintain linearity at different concentration. In precision study, the cumulative %RSD of silodosin was 0.88. The percent recovery of the silodosin at different concentration was within the (98.0-102.0) %. When the column temperature was increased/decreased by 3°C from the real and the flow rate was increased/decreased by 0.2mL/min from the actual rate, the system suitability resolution was still within the acceptable range. The standard and sample solution were stable after 24 hours at both room temperature and 5°C temperature. For the analysis of silodosin in pharmaceutical goods, the confirmed HPLC method may be a workable analytical approach.
KEYWORDS: Silodosin, HPLC, Benign prostatic hyperplasia, Chromatography, Validation.
INTRODUCTION:
Beyond the age of 40, Benign Prostatic Hyperplasia (BPH) affects many males. By the age of 60, BPH affects around 50% of men; by the age of 85, it affects 90% of men1-3. 50% of BPH patients report having symptoms of the urinary tract (lower) infection2.
One of the BPH therapies for mild Lower Urinary Tract Syndrome (LUTS) is watchful waiting, whereas moderate to severe LUTS is treated with pharmacologic therapy (such as 5-reductase inhibitors or 1-adrenergic receptor antagonists), and severe LUTS is treated with surgery1.
Alpha-blockers are still the treatment for LUTS in BPH. The FDA has authorized the new 1-Alpha-blocker silodosin since October 2008 at the recommended dose of 8 mg once daily. Clinical studies have shown that this selective 1-Alpha-blocker is much more desirable and effectivefor LUTS in BPH. Silodosin exhibits exceptional early efficacy and is as least as effective as other 1-blockers4. It has been demonstrated that silodosin inhibits the contraction of the prostate smooth muscle and has a better selectivity for the dominant subtype of the Alpha-1-Adrenergic receptor in prostate tissues. As a result, it is frequently utilized in therapeutic settings to treat BHP5-7.
The goal of the research was to create a HPLC method that would be cost effective, short process, and appropriate for daily laboratory analysis of silodosin found in capsule formulation.Both qualitative and quantitative testing of dosage forms is required for pharmaceutical products quality. To ensure the stability and safety of the formulation products, quantitative testing must be carried out during the course of its lifecycle.
According to a literature study, many methods have been employed to determine silodosin. Some of the methods were described in HPLC8-15, Ultra Performance Liquid Chromatography (UPLC)16, UV-Vis spectrophotometry17-19, spectrofluorimetric20, and electrochemical methods21,22.
The most convenient instrumental analytical method for determination of active pharmaceutical ingredients in pharmaceutical products is RP-HPLC method23-35.
An effective, stability focused HPLC method was developed, but the method was notreported all validation parameters, described in ICH guidelines8. Two Liquid Chromatography-Mass Spectroscopy (LC-MS) methods were developed and validated successfully for silodosin detection in human plasma, not in capsule formulation9,12. By using reverse phase HPLC techniques with DAD and UV detectors, silodosin was separated successfully and the methods was validated successfully within the short retention time, but these methods were applied in silodosin with others active ingredients13-15. Another recently published study validated Quality by Design (QbD) based method for silodosin in tablet dosage form, but method was used florescence detector, which is not available in all laboratory11.
According to the literature study, there isn't a single analytical easy, precise, and robust technique forevaluating silodosin in capsuleformulation both qualitatively and quantitatively.
The intention of this work was to build a simple, repeatable, linear, robust, and accurate single analytical approach for the simultaneous measurement of silodosin in capsule formulation. According to a thorough examination of the literature, for the rapid and simple detection of silodosin in capsule formulations, the current method is the first robust method ever created.
MATERIALS AND METHOD:
Instrument:
Agilent HPLC 1260 Infinity series for the entire study was used along with DAD detector. The software, empower was used for the processing of Chromatogram. Chromatographic Column for separation the peak was used Zorbaz brand. Ultrasonic Bath was Power sonic 620, Hwashin, Korea brand. Analytical balance for weighed the chemicals and reagents was Mettler Toledo, Switzerland brand. pH meter was Mettler Toledo, Switzerland brand. All the Instruments were calibrated with appropriate instruction before use.
Chemical and Reagent:
Reagents were used from different reliable resources. Silodosin was purchased from Lewens Lans pvt. Ltd. Gujarat, India. Finished product, Silodosin capsule was purchased from local pharmacy. Analytical grade Reagents, Acetonitrile from RCI labs can ltd, Thailand and Methanol from Supelco Inc. Ammonium Acetate from Samchun, was used throughout the study. Milli-Q was used to obtain purified water. Mannitol (DC grade), Pregelatinised starch (starch 1500), Sodium Lauryl Sulfate, and Magnesium Stearate were used as excipients in capsule formulation.
Solution Preparation:
Mobile Phase preparation: Taken 300mL of Bufferand 300mL of Acetonitrile into 1L of beaker. Added 400mL of Methanol, sonicated for 5minutes, after that filtered the solution through 0.22µ membrane filter and degassed 10minutes prior to use. Mobile Phase was used as diluent.
Preparation of Buffer: Weighed1.156g of Ammonium Acetate into 1L of beaker containing 1L of water and sonicated for 5 minutes then filtered through 0.22µm membrane filter and degassed 10minutes before use.
Standard solution: 0.025mg/mL silodosin solution was prepared by transferring 25.0mg of silodosin working standard into a 100mL volumetric flask. Pipetted out 5 mL of this solution in 50mL another volumetric flask. Filtered the preparation using 0.22µm PVDF syringe filter before injection.
Sample solution: 0.025mg/mL silodosin sample solution was prepared by transferring 781.25 mg crushed powder of capsule (equivalent to 25mg of Silodosin) to a 100 mL volumetric flask. Accurately pipetted out 5mL of solution in 50mL of volumetric flask. Filtered the preparation using 0.22µm PVDF syringe filter before injection.
Chromatographic Condition:
The separation was done with Cyano; 4.6mm × 25cm; 5 mm, packing USP Octadecyl Silane columnand peak was detected at 270nm wavelength. The flow rate was 1.0 mL/min, Injection volume was 20µl, and column oven temperature was 40°C.
Analytical Method Validation:
The method was validated according to ICH guideline ICH-Q2 (R1)36 and United States of Pharmacopeia (USP) general chapter 122537 with respect to system suitability, specificity, LOD/LOQ, linearity, precision, accuracy, robustness and solution stability.
System suitability study:
To check the system for suitability, the Relative Standard Deviation (RSD) from five injections of standard solution, tailing factor and USP theoretical plate count were checked. The %RSD of Silodosin must be within the 2.0, Tailing factors not more than 2.0, and USP theoretical plate must be not less than 2000.
Specificity:
To check the interference of diluent and placebo with silodosin, specificity study was done. The separate 0.025 mg/mL silodosin standard/sample solution andplacebo (mixture of excipients)were prepared and injected individually. The chromatogram produced from the diluent shouldn't have any interfering peaks at the retention time that corresponds to the peak for silodosin. The retention time of peak due to silodosin in sample solution shall be similar (±2%) to that of standard. Peak purity should pass i.e.; Purity angle should be less than purity threshold.
LOD/LOQ:
Limit of Detection (LOD) and Limit of Quantification (LOQ) were determined by using the signal to noise (S/N) ratio method. Prepared silodosin standard solution at different low-level of concentration as much as possible and injected. Determined the LOD and LOQ using equation: LOD = 3.3 σ/ s and LOQ = 10 σ/ s. [Where, σ = the standard deviation of the response and S = the slope of the calibration curve (estimated from the linearity curve)].
After determination of LOD and LOQ Concentration, LOD and LOQ solution were prepared and injected. In case of LOQ, S/N ratio of silodosin peaks must be greater than or equal to 10. The peak at LOD level should be detectable and S/N ratio must be more than 3.
Linearity:
Linearity of silodosin was demonstrated by injecting silodosin standard solutionin concentration from LOQ concentration level to 200% of 100% concentration of test solution (0.025mg/mL). The correlation coefficient (R2) was calculated by using calibration curve. The R2 must not be less than 0.90.
Preparation of linearity stock solution: Transferred 25.00 mg of the Silodosin standard into a 100mL volumetric flask. Then, added 70mL of diluent, sonicate to dissolve it, and bring the volume up to the desired level with the diluent. Finally, mixed well. The concentration of silodosin was 0.25mg/mL. Table 1 represents the concentration of linearity solution.
Table 1. Preparation of linearity solutions (LOQ to 200%)
|
Conc. (%) |
Volume taken from Linearity stock solution (mL) |
Total volume with diluent (mL) |
Concentration (mg/mL) |
|
Silodosin |
|||
|
5% |
1 |
250 |
0.001 |
|
10% |
2 |
200 |
0.003 |
|
30% |
3 |
100 |
0.008 |
|
50 % |
5 |
100 |
0.013 |
|
80% |
4 |
50 |
0.020 |
|
100 % |
5 |
50 |
0.025 |
|
120 % |
6 |
50 |
0.030 |
|
150 % |
15 |
100 |
0.038 |
|
200 % |
10 |
50 |
0.051 |
Precision:
Method Precision Study: Assess the repeatability by performing six replicate measurements (n=6) of the assay of Silodosin 4 mg Capsule as per test method at 100% of test concentration (0.025mg/mL) and calculating the %RSD of the assay of Silodosin. The assay of Silodosin must be within 98.0% - 102.0% of label claim and %RSD of assay of Silodosin must not be more than 2.0.
Intermediate Precision Study:
The intermediate precision was done by another analyst with different HPLC at different day. Cumulative %RSD of each injection from 12 result of injections (method precision and intermediate precision) should be not more than 10.0%.
Accuracy:
To established the accuracy of the analytical method, prepared spiked sample solution by spiking silodosin API and Placebo (Mixture of excipients) at LOQ, 50%, 100% and 150% of test concentration (0.025mg/mL). The recovery (%) should be not less than 98.0% and should be not more than 102.0%.
Robustness:
Prepared the standard solution at test concentration (0.025mg/mL). Inject the prepared solution using different chromatographic conditions as shown below:
1. By changing the Column Temperature by +3°C from the actual column temperature (40°C).
2. By changing the flow rate by ± 0.2 mL/min from the actual flow rate (1.0 mL/min).
3. By changing the Mobile phase composition, buffer and organic ratio by ±2%from the actual ratio.
The %RSD for replicate injection must not be more than 2.0, theoretical plates more than 2000, and tailing must be with in 2.0.
Solution Stability:
To conducted solution stability, the
standard solution, sample solution and spiked sample solution were kept in room
temperature and refrigerator (5
) for 24 hours. After 24 hours, all
solution injected and compared with the initial results.
RESULTS AND DISCUSSION:
System Suitability:
The column efficiency determined from Silodosin peak was found 9224 theoretical plates, tailing factor of the Silodosinpeakwas found 0.05, and the %RSD of five standard solution was found 1.13%, indicate that the system is suitable for analysis.
Specificity study:
It was observed that after all specific injection, silodosin gave same response in sample (RT-3.992 minutes) and standard (RT-3.992 minutes) with respect to retention time (RT) and no interference due to diluent and placebo. The peak purity angle more than the purity threshold in all the single injection.Therefore, the peak response of method is considered as specific. Specificity study data are presented in Table 2.
Table 2. Specificity Study with peak purity angle and threshold
|
Name of the Solution |
Name of Peak |
Retention time (min) |
Diluent interference |
Purity Angle |
Purity Threshold |
|
Diluent |
- |
- |
Nil |
- |
- |
|
Placebo |
- |
- |
- |
- |
- |
|
Standard Solution |
Silodosin |
3.992 |
- |
3.86 |
4.25 |
|
Sample Solution |
Silodosin |
3.992 |
- |
3.24 |
5.11 |
HPLC Chromatogram of standard, sample, diluent, and placebo are given in the Figure 1-4.
Figure 1. Chromatogram of Standard
Figure 2. Chromatogram of Sample
Figure 3. Chromatogram of Diluent
Figure 4. Chromatogram of Placebo
LOD/LOQ:
The value of LOD was found 0.00036 mg/mL. The LOQ value was found 0.00109 mg/mL.S/N ratio of silodosin peaks was within the acceptance limit. The S/N ratio were found greater than 3 in case of LOD, and greater than 10 in case of LOQ. The peak at LOD level was detectable in all six replicates.
Linearity:
The correlation coefficient of silodosin was found 1.00.
Figure 3: Linearity graph for Silodosin (Concentration vs. Area)
The results were well within acceptance limit. Thus, the method is considered as linear. Linearity data are presented in Figure 3.
Precision:
Method Precision/Repeatability Study: % RSD of silodosin was 0.52% in six sample solutions. Assay result in sample solution was found 98.92% of the level claim of silodosin, the method meets the requirement of method precision.
Intermediate Precision/Repeatability Study:
%RSD of silodosin was 0.14% in six sample solutions. Assay result in sample solution was found 98.24% of the level claim of silodosin, the method meets the requirement of intermediate precision.
Cumulative % RSD of silodosin was 0.88% in 12 injections (method precision and intermediate precision). Hence, the method meets the requirement of Precision study/Ruggedness.
Accuracy:
The accuracy in terms of percent recovery of four different concentration (s) were well enough within the acceptance limit (98.00-102.00) %. The %RSD was below 2.0. Hence, the method meets the requirement of accuracy. The Accuracy study data are presented in Table 3.
Table 3. Silodosin Recovered in accuracy study
|
Sample Name |
Amount of silodosin API taken (mg) |
Area of Silodosin |
Amount recovered (mg) |
% Recovery |
|
Accuracy LOQ -1 |
1.01 |
20643 |
0.99 |
98.46 |
|
Accuracy LOQ -2 |
1.04 |
21323 |
1.03 |
98.77 |
|
Accuracy LOQ -3 |
1.03 |
21324 |
1.30 |
99.73 |
|
Accuracy 80% -1 |
20.11 |
417248 |
20.10 |
99.95 |
|
Accuracy 80% -2 |
20.13 |
414962 |
19.99 |
99.31 |
|
Accuracy 80% -3 |
20.05 |
411537 |
19.83 |
98.88 |
|
Accuracy 100% -1 |
25.20 |
516264 |
24.87 |
98.69 |
|
Accuracy 100% -2 |
25.23 |
521925 |
25.14 |
99.66 |
|
Accuracy 100% -3 |
25.21 |
521035 |
25.10 |
99.57 |
|
Accuracy 120% -1 |
30.08 |
617353 |
29.74 |
98.87 |
|
Accuracy 120% -2 |
30.04 |
616947 |
29.72 |
98.94 |
|
Accuracy 120% -3 |
30.11 |
612812 |
29.52 |
98.05 |
|
|
|
Minimum |
|
98.05 |
|
|
|
Maximum |
|
99.95 |
|
|
|
Average |
|
99.1 |
|
|
|
%RSD |
|
0.59 |
Robustness Study:
The robustness study was satisfactory with column oven temperature, flow rate, and mobile phase composition variation. The %RSD of silodosin standard was found are given in Table 4.
Table 4. %RSD at different condition from actual condition in robustness study
|
Condition |
%RSD |
|
Column oven temperature at 37°C |
0.43 |
|
Column oven temperature at 43°C |
0.56 |
|
Flow rate at 0.8 mL/min |
0.85 |
|
Flow rate at 1.2 mL/min |
0.77 |
|
Mobile Phase ratio, Buffer: Acetonitrile: Methanol (29.40: 30.30:40.30) |
0.55 |
|
Mobile Phase ratio, Buffer: Acetonitrile: Methanol (30.60:29.70:39.70) |
0.53 |
Solution Stability Study:
The solution of standard, sample, and spiked sample was stable up to 24 hours as they provided same response as initially provided even if in the room temperature and refrigerator (5°C). The differences from initial to 24 hours result are presented in Table 5.
Table 5. % Difference of initial fresh solution and solution after 24 hours
|
Sample Name |
% Difference at room temperature |
% Difference at 5°C temperature |
|
Standard Solution |
0.6 |
0.5 |
|
Sample Solution |
0.4 |
0.5 |
|
Spiked Sample Solution |
0.5 |
0.4 |
CONCLUSION:
This work was aimed to developed short run time, simple, easy to conduct, and cost-effective analytical HPLC method for determination of silodosin in capsule dosage form. In pharmaceutical manufacturing site, finished products required quantification before marketed. The literature review concluded that there are no easy, robust validated analytical methods for silodosin detection with less cost. Based on the findings, the placebo interference, linearity, accuracy, precision, robustness, and solution stabilitymet the criteria. The logical conclusion is that silodosin can be detected using this analytical approach and this method is convenient enough for daily analysis.
ACKNOWLEDGMENTS:
We are thankful to the respectful teacher of Department of Pharmacy, World University of Bangladesh for providing facilities to conduct this work. All co-authors are contributed equally in writing the manuscripts and compared with published data.
CONFLICTS OF INTEREST:
The authors declare that there are no conflicts of interest regarding the publication of this article.
FUNDING:
This work was conducted by self-fund.
REFERENCES:
1. Committee AUA P G. AUA guideline on management of benign prostatic hyperplasia (2003) Chapter 1: Diagnosis and treatment recommendations. The Journal of Urology. 2003; Aug; 170(2 pt 1): 530-547. doi.org/10.1097/01.ju.0000078083.38675.79
2. Wasserman N F. Benign prostatic hyperplasia: A review and ultrasound classification. Radiologic Clinics of North America. 2006; Sep; 44(5): 689-710.doi.org/10.1016/j.rcl.2006.07.005.
3. Thorpe A, Neal P D. Benign prostatic hyperplasia. The Lancet. 2003; 362(9366): 1359-1367. doi.org/10.1016/S0140-6736(03)13073-5
4. Rossi M, Roumeguere T. Silodosin if the treatment of benign prostatic hyperplasia. Drug design, development and therapy. 2010; 4: 291-297. doi.org/10.2147/DDDT.S10428
5. Marks L S, Gittelman M C, Hill L A, Volinn W, Hoel G. Silodosin in the Treatment of the Signs and Symptoms of Benign Prostatic Hyperplasia: A 9-Month, Open-label Extension Study. Urology. 2009; 74: 1318–1322. doi: 10.1016/j.urology.2009.06.072
6. Roehrborn C G, Cruz F, Fusco F. α1-Blockers in Men with Lower Urinary Tract Symptoms Suggestive of Benign Prostatic Obstruction: Is Silodosin Different? Advances in Therapy. 2017; 33: 2110–2121. doi:10.1007/s12325-016-0423-5
7. Yoshida M, Kudoh J, Homma Y, Kawabe K. Safety and efficacy of silodosin for the treatment of benign prostatic hyperplasia. Clinical Interventions in Aging. 2011; 6: 161–172. doi: 10.2147/cia.s13803
8. Er E, Erk N. An effective and sensitive stability-indicating chromatographic approach based on HPLC for silodosin assay. Journal of Analytical Science and Technology. 2016; 7:20. doi.org/10.1186/s40543-016-0100-y
9. Zhao X, Liu Y, Xu J, Zhang D, Zhou Y, Gu J, Cui Y. Determination of silodosin in human plasma by liquid chromatography-tandem mass spectrometry. Journal of Chromatography B. 2009; 877(29): 3724–3728. doi.org/10.1016/j.jchromb.2009.08.025
10. Harischandran S, Shankar L R, Raju R, Shibi A, Sayana P S. Method development and validation RP-HPLC method for assay of silodosin in pharmaceutical dosage form. International Journal of Pharmaceutical Sciences and Research. 2012; 1(4): 141–145.
11. Yaman M E, Akman T C. Optimization of HPLC-FLD Conditions Using Analytical Quality by Design Approach for Quantification of Silodosin in Pharmaceutical Dosage Form. Erzincan University Journal of Science and Technology. 2021; 14 (3): 993-1005. doi.org/10.18185/erzifbed.955967
12. Yadlapalli S S R, Katari N K, Manabolu S B, Ravi V B. Simultaneous estimation of silodosin and silodosin β-Dglucuronide in human plasma using LC-MS/MS for a pharmacokinetic study. Journal of Taibah University for Science. 2018; 12 (4): 393-400. doi.org/10.1080/16583655.2018.1490562
13. Nataraj KS, Srinivasa R A, Harshitha S, Pravallika K. Analytical Method Development and Validation for the Estimation of Silodosin and Dutasteride in Capsule Dosage Form by RP-HPLC Method. Journal of Drug Metabolism & Toxicology. 2020; 11(2): 245. 10.35248/2157-7609.20.11.245
14. Shah H P, Khandar A, Deshpande S, Bagade S. Novel RP-HPLC method for simultaneous estimation of silodosin and dutasteride in multiunit solid dosage form. Research Journal of Pharmaceutical, Biological and Chemical Sciences. 2014; 5(2): 801-810.
15. Gupta A, Mishra S K. A Novel Analytical Method for Simultaneous Quantification of Silodosin and Tadalafil by RP-HPLC. Journal of Pharmaceutical Research International. 2021; 33(39B): 193-202. doi.org/10.9734/jpri/2021/v33i39B32195
16. Shaik J V, Saladi S, Sait S S. Development of Stability-Indicating UHPLC Method for the Quantitative Determination of Silodosin and Its Related Substances. Journal of Chromatographic Science. 2014; 52(7): 646–653. doi.org/10.1093/chromsci/bmt094
17. Ahmed S, Abdelkawy M, Mohamed T A, Nasser N. Validated stability-indicating spectrophotometric methods for the determination of Silodosin in the presence of its degradation products.Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2018; 202: 131–145. doi.org/10.1016/j.saa.2018.04.046
18. Mounika C H, Umadevi N, Sudheerbabu I. New visible spectrophotometric for the estimation of silodosin in pharmaceutical formulations. International Journal of Research in Pharmacy and Chemistry. 2013; 3(3): 595–597.
19. Aneesh T P, Aiyalu R. Method development and validation for the estimation of silodosin in bulk and pharmaceutical dosage forms using UV-VIS spectrophotometry. Asian Journal of Pharmaceutical and Clinical Research. 2012; 5(4): 4–6.
20. Bhamre P, Rajput S J. Spectrofluorimetric method for the determination of silodosin in bulk and pharmaceutical dosage form. Indo American Journal of Pharmaceutical Research. 2014; 4(10): 5106–5110.
21. Toker B, Er E, Erk N. Development of voltammetric techniques for the determination of silodosin in pharmaceutical formulation at glassy carbon electrode. Revue Roumaine de Chimie. 2014; 59(5): 311–316.
22. Er E, Celikkan H, Erk, N. An ultra-sensitive 2D electrochemical sensor based on a PtNPs@graphene/Nafion nanocomposite for determination of α1-AR antagonist silodosin in human plasma. Analytical Methods. 2017; 9: 3782–3789.
23. Mahmoud M. Sebaiy, Abdullah A. El-Shanawany, Sobhy M. El-Adl, Lobna M. Abdel-Aziz, Hisham A. Hashem. Rapid RP-HPLC Method for Simultaneous Estimation of Norfloxacin and Tinidazole in Tablet Dosage Form. Asian J. Pharm. Ana. 2011; 1(4): Oct. - Dec. 79-84.
24. L. Satyanarayana, S.V. Naidu, M. Narasimha Rao, Reddy Suma Latha. The Estimation of Nilotinib in Capsule dosage form by RP-HPLC. Asian J. Pharm. Ana. 2011; 1(4): 100-102.
25. Sharmin Reza Chowdhury, Mahfuza Maleque, Mahbubul Hoque Shihan. Development and Validation of a Simple RP-HPLC Method for Determination of Caffeine in Pharmaceutical Dosage Forms. Asian J. Pharm. Ana. 2012; 2(1): 1-4.
26. P. Janaki Pathi, N. Appala Raju. The Estimation of Epalrestat in Tablet Dosage Form by RP-HPLC. Asian J. Pharm. Ana. 2012; 2(2): 49-51.
27. Y Rajendraprasad, KK Rajasekhar, V Shankarananth, S Bharath kumar, G Dileep kumar, T Sreelekha. RP-HPLC Method for the Estimation of Domperidone in Bulk and Pharmaceutical Formulations. Asian J. Research Chem. 2009; 2(4): 561-564.
28. K Ananda kumar, G Kumaraswamy, T Ayyappan, ASK Sankar, D Nagavalli. Validated RP-HPLC Method for the Estimation of Eszopiclone in Bulk and Tablet Dosage Form. Asian J. Research Chem. 2010; 3(1): 63-66.
29. Antakli S, Alahmad A, Badingki H. Simultaneous Determination of Sodium Benzoate and Potassium Sorbate Preservatives in Foodstuffs Using High-Performance Liquid Chromatography. Asian J. Research Chem. 2010; 3(1): 178-182.
30. Amrish Sharma, Mukul Tailang, Bhaskar Gupta, Ashish Acharya. RP HPLC Determination of Amlodipine in Its Pharmaceutical Dosage Forms. Asian J. Research Chem. 2010; 3(1): 129-131.
31. Prabhu S, Vijay Amirtharaj R, Senthilkumar N. Simultaneous RP-HPLC Method Development and Validation of Cefixime and Ofloxacin in Tablet Dosage Form. Asian J. Research Chem. 2010; 3(2): 367-369.
32. T.M. Kalyankar, R.B. Kakde. Reversed-Phase Liquid Chromatographic Method for Simultaneous Determination of Metformin and Repaglinide in Pharmaceutical Preparations. Research J. Pharma. Dosage Forms and Tech. 2011; 3(5): 233-237.
33. Shalini Pant, Krishan Pal. Development and Validation of a Simultaneous HPLC Method for Assay and Dissolution of Bisoprolol fumarate and Amlodipine besylate in Pharmaceutical Dosage. Research J. Pharma. Dosage Forms and Tech. 2012; 4(1): 62-66,
34. N Appala Raju, Shabana Begum. Simultaneous RP-HPLC Method for the Estimation of the Emtricitabine, Tenofovir Disoproxil Fumerate and Efavirenz in Tablet Dosage Forms. Research J. Pharm. and Tech. 2008; 1(4): 522-525.
35. Akash Shelke, Someshwar Mankar, Mahesh Kolhe. Development and Validation of RP-HPLC Method for estimation of Secnidazole in API and Pharmaceutical Dosage Form. Research Journal of Science and Technology. 2021; 13(2): 100-4.
36. International Conference on Harmonization (ICH) of Technical Requirements for Registration of Pharmaceuticals for Human Use, Topic Q2 (R1). Validation of Analytical Procedures: Text and Methodology. International Conference on Harmonization (ICH), Geneva, Switzerland. 2005 November.
37. U.S. Pharmacopeial Convention. The U.S. Pharmacopoeia (USP 43), National Formulary (NF 38) General Chapters: 1225. Validation of Compendial Procedures. Rockville. 2020.
Received on 07.01.2024 Modified on 06.05.2024
Accepted on 13.07.2024 © RJPT All right reserved
Research J. Pharm. and Tech 2024; 17(9):4311-4317.