Author(s): Bagus Dwi Irianto Widodo, Yetti Hernaningsih, Hartono Kahar

Email(s): yetti-h@fk.unair.ac.id

DOI: 10.52711/0974-360X.2025.00880   

Address: Bagus Dwi Irianto Widodo1, Yetti Hernaningsih2,3*, Hartono Kahar2,3
1Clinical Pathology Residency Program, Department of Clinical Pathology, Faculty of Medicine, Universitas Airlangga - Dr. Soetomo General Academic Hospital, Surabaya, Indonesia.
2Department of Clinical Pathology, Faculty of Medicine, Universitas Airlangga, Surabaya, Indonesia.
3Dr. Soetomo General Academic Hospital, Surabaya, Indonesia.
*Corresponding Author

Published In:   Volume - 18,      Issue - 12,     Year - 2025


ABSTRACT:
HumaCount 5D CRP is an instrument for hematology examination that comprises hemoglobin, leukocytes, erythrocytes, hematocrit, and platelet examinations. Sysmex XN-1000 is one of the current established instruments for hematology examination. Comparison of the results between new and established instruments can be done by comparing each parameter in all result categories (low, normal, high). This study aims to analyze and compare the results of hematology examinations between the HumaCount 5D CRP and one of the current established instruments (the Sysmex XN-1000) on 5 parameters of hemoglobin, leukocytes, erythrocytes, hematocrit, and platelets. Comparative tests were conducted on 172 samples using HumaCount 5D CRP and Sysmex XN-1000. The result of each examined parameter was categorized into low, normal, and high. SPSS was used to analyze the data on the entire result, and each group of parameters using the normality test, and further was adjusted using parametric and non-parametric tests. The examination of the whole samples using the One-Way ANOVA test showed that the results were hemoglobin (p=0.818), leukocytes (p=0.325), erythrocytes (p=0.626), hematocrit (p=0.000), and platelets (p=0.394). Testing each parameter group, parametric and non-parametric tests were used to adjust the results of the normality test and the results were leukocytes (low p=0.109, normal p=0.010, high p=0.020), erythrocytes (low p=0.015, normal p=0.825), hemoglobin (low p=0.167, normal p=0.898), hematocrit (low p=0.000, normal p=0.000), and platelets (low p=0.472, normal p=0.088, high p=0.695). The overall test results showed that the HumaCount 5D CRP instrument has quite good results. The HumaCount 5D CRP is a fairly good hematology analyzer instrument and can be used for clinical services. However, it is necessary to be aware of the hematocrit result if there is a discordance with the patient's clinical condition.


Cite this article:
Bagus Dwi Irianto Widodo, Yetti Hernaningsih, Hartono Kahar. Compatibility of Complete Blood Count Values between HumaCount 5D CRP and Sysmex XN - 1000. Research Journal Pharmacy and Technology. 2025;18(12):6084-0. doi: 10.52711/0974-360X.2025.00880

Cite(Electronic):
Bagus Dwi Irianto Widodo, Yetti Hernaningsih, Hartono Kahar. Compatibility of Complete Blood Count Values between HumaCount 5D CRP and Sysmex XN - 1000. Research Journal Pharmacy and Technology. 2025;18(12):6084-0. doi: 10.52711/0974-360X.2025.00880   Available on: https://www.rjptonline.org/AbstractView.aspx?PID=2025-18-12-70


REFERENCES: 
1.    Indonesian Ministry of Health. Indonesia Health Profile 2021. 1st ed.Vol. 1. Indonesian Ministry of Health, Jakarta. 2022.
2.    Putra AAMS. Hernaningsih Y. Comparison of Complete Blood Count Parameters using EDTA, Sodium Citrate, and Heparin Anticoagulants. Research J. Pharm. and Tech. 2022; 15: 4687-1. doi: 10.52711/0974-360X.2022.00786.
3.    Nima RS. Ablaa HA. Mohammed HJ. The Relationship Between Blood Counts and Serum  Levels of zinc and Copper in Children with β-thalassemia Living in Najaf Governorate, in  Iraq. Research J. Pharm. and Tech. 2018; 11: 4640-4644. doi: 10.5958/0974-360X.2018.00848.X.
4.    Bhola RK. Fudaly C. Rastogi S. A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analysers. Indian J Hematol Blood Transfus. 2024; 40: 303–314. https://doi.org/10.1007/s12288-023-01687-6.
5.    Permatasari CA. Hajat A. Sedana MP. Diagnostic Testing of Hematology Analyzer Equipped with WPC Channel Scattergram in Determining Strains of Hematological Malignancies. Research J. Pharm. and Tech. 2022; 15: 4005-1. doi: 10.52711/0974-360X.2022.00671.
6.    Putra IGPAFS. Prihatiningsih D. Subhaktiyasa PG. Analysis of Internal Quality Assurance of Platelet Examination in the Clinical Laboratory of Abiansemal I Primary Health. Jurnal Analis Laboratorium Medik. 2020; 5(2): 28–34. https://doi.org/10.51544/jalm.v5i2.2152.
7.    Bain BJ, Bates I, Laffan MA. Dacie and Lewis Practical Haematology. 12th ed. Vol. 1. Elsevier, London. 2017.
8.    Sysmex Company. PPT: Sample Analysis. Sysmex Asia Pasific Pte. Ltd, 2011.
9.    Yasmeen R. Reddy NK. Souza MGD. Swathi B. Waheed AA. Chinmayee K. Reddy JV. Study of Anticoagulants low molecular weight heparin and Unfractionated heparin in the management of Non-St elevation Myocardial Infarction. Research J. Pharm. and Tech. 2020; 13: 3151-3155. DOI: 10.5958/0974-360X.2020.00557.0
10.    Hussein TA. Saihood AS. Rayshan AR. Molecular Identification and Characterization in blood samples of patients with progressive multifocal leukoencephalopathy and human kidney cell line culture of John Cunningham Virus (JCV) in Iraq. Research J. Pharm. and Tech. 2018; 11: 3392-3398. DOI: 10.5958/0974-360X.2018.00625.X
11.    Victor ON. Tufon EN. Osmond EC. Effect of Vitamin E-Stored Experimental Blood Samples on Malondialdehyde and Protein Concentration. Research J. Pharm. and Tech. 2011; 4: 1785-1786.
12.    Human Gesellschaft für Biochemica und Diagnostica mbH. HumaCount 5DCRP: User Manual. Rev. 04/2023-02. Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden, Germany. 2023.
13.    Varanita SV. Kahar H. Fitriah M. Compatibility of Two EDTA Tubes for Hemoglobin (Hb) Hematocrit (Hct) Tests with Aptus Equipment. Research J. Pharm. and Tech. 2022; 15: 4447-0. doi: 10.52711/0974-360X.2022.00745.
14.    Subbalakshmi NK. Sunandha S. Influence of Perceived Stress on Hemoglobin Concentration. Research J. Pharm. and Tech. 2017; 10: 61-64. DOI: 10.5958/0974-360X.2017.00015.4.
15.    Ahmad RI. Al-Barry HA. Salloum SI. Fetal Hemoglobin (HbF) levels in Sickle cell anemia patients in Lattakia, Syria. Research J. Pharm. and Tech. 2020; 13: 3246-3248. DOI: 10.5958/0974-360X.2020.00575.2.
16.    Brundha MP. Pathmashri VP. Sundari S. Quantitative Changes of Red Blood cells in Cancer Patients under Palliative Radiotherapy-A Retrospective Study. Research J. Pharm. and Tech. 2019; 12(2):687-692. DOI: 10.5958/0974-360X.2019.00122.7.
17.    Wang J. Zhao S. Su Z. Liu X. Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000. J Clin Lab Anal. 2019; 33: e22955. https://doi.org/10.1002/jcla.22955.

Recomonded Articles:

Research Journal of Pharmacy and Technology (RJPT) is an international, peer-reviewed, multidisciplinary journal.... Read more >>>

RNI: CHHENG00387/33/1/2008-TC                     
DOI: 10.52711/0974-360X 

1.3
2021CiteScore
 
56th percentile
Powered by  Scopus


SCImago Journal & Country Rank

Journal Policies & Information


Recent Articles




Tags


Not Available